Siemens Healthcare Diagnostics, Inc.: ADVIA Centaur Erythropoietin (EPO) Assay (100 Test)

Hazard

Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.

Details from the source

Reason for recall: Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots. Distribution: Worldwide - US Nationwide distribution in the states of AL, CA, IA, IL, MI, MN, NC, NE, NV, NY, OH and TX; The countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, P.R. China, Pakistan, Poland, Portugal, Republic of South Korea, Romania, Quantity: 264 units (6 US, 258 OUS) Lot/code info: Siemens Material Number: 10995096; UDI/DI: 00630414010380; Kit Lots: 53984040, 55443042.

UPC codes

00630414010380

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify