There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.
Reason for recall: There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy. Distribution: Distribution list not yet provided. Quantity: 391 total IABPs Lot/code info: Model Number: 0998-00-0800-75, UDI-DI: 10607567112312; Model Number: 0998-00-0800-83, UDI-DI: 10607567108407; Model Number: 0998-00-0800-85, UDI-DI: 10607567113449; Serial Numbers: CA237026J3 CH231276C6
10607567112312, 10607567108407, 10607567113449
Official record: FDA (Device enforcement) ↗