← Search recalls
FDA (Device enforcement)Recalled 2023-02-13class i

Abbott Diabetes Care, Inc.: FreeStyle Libre Reader, REF: 71525-01, 71701-01, 71739-01 a component of the FreeStyle Libre Flash Glucose Monitoring System

Hazard

Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.

Details from the source

Reason for recall: Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use. Distribution: U.S. Nationwide. Quantity: 336,085 Lot/code info: REF/UDI-DI/Lot: 71525-01/357599000219/02Y702D, 01Y699D, 02Y699D, 01Y698D, 02Y697D, 04Y690D, 03Y690D, 01Y683D, 02Y683D, 01Y674D, 02Y672D, 01Y669D, 02Y666D, 01Y659D, 02Y658D, 02Y656D, 04Y655D, 01Y655D, 01Y652D, 02Y653D, 03Y653D, 03Y651D, 01Y641D, 02Y638D, 01Y637D, 02Y635D, 01Y629D, 03Y628D, 01Y628D, 02Y628D, 02Y621D, 02Y622D, 02Y620D, 02Y608D, 04Y608D, 01Y608D, 01Y603D, 03Y599D, 04Y600D, 02Y596D, 06Y596D, 03Y596D, 07Y547D, 07Y546D, 08Y546D, 05Y540D, 06Y538D, 05Y537D, 03Y536D, 04Y523D, 02Y522D, 02Y519D, 02Y512D, 01Y512D, 01Y511D, 02Y510D, 01Y505D, 02Y504D, 04Y355D, 02Y355D, 01Y352D, 01Y349D, 03Y348D, 01Y348D, 02Y348D, 01Y339D, 01Y335D, 02Y334D, 02Y335D, 03Y314D, 01Y313D, 02Y312D, 01Y312D, 02Y311D, 01Y311D, 02Y310D, 04Y310D, 02Y307D, 03Y307D, 01Y304D, 02Y303D, 02Y304D; 71701-01/357599000226/01Y702D, 01Y694D, 03Y693D, 02Y686D, 03Y683D, 02Y574D, 01Y563D, 02Y563D, 01Y564D, 03Y562D, 01Y552D, 02Y552D, 06Y550D, 05Y539D, 05Y538D, 06Y537D, 01Y536D, 01Y532D, 02Y531D, 03Y522D, 02Y518D, 02Y511D, 01Y508D, 02Y505D, 04Y356D, 01Y356D, 02Y356D, 01Y321D, 01Y320D, 02Y320D, 01Y317D, 02Y317D; 71739-01/357599000240/03Y754D, 03Y734D, 01Y705D, 01Y671D

UPC codes

357599000219, 357599000226, 357599000240

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify