The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.
Reason for recall: The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications. Distribution: US Nationwide distribution in the state of IL. Quantity: 50000 ea/ 500 cases Lot/code info: UDI (01)00810050852210(10)C2O24002(11)240626(17)290625 Lot C2O24002
00810050852210
Official record: FDA (Device enforcement) ↗