← Search recalls
FDA (Device enforcement)Recalled 2024-02-12class i

ARROW INTERNATIONAL Inc.: ARROW QuickFlash Radial Artery Catherization (wire, guide, catheter), Part Numbers: a) REF AK-04220; b) REF ASK-04220-FMH1; c) REF ASK-04220-HHC; d) REF ASK-04220-KSP; e) REF

Hazard

Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use. The possible immediate health consequences of the component issue are arterial vasospasm and vessel injury arising from multiple arterial punctures with repeated attempts.

Details from the source

Reason for recall: Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use. The possible immediate health consequences of the component issue are arterial vasospasm and vessel injury arising from multiple arterial punctures with repeated attempts. Distribution: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: AE, AR, AT, AU, BE, BH, BR, CA, CN, CO, CY, DE, EC, EE, ES, FR, GR, HK, HR, IT, KW, LB, LU, MX, MY, NL, NZ, PH, PL, QA, RU, SA, SE, SG, SI, TH, TW, and UK. Quantity: 1,160,470 devices Lot/code info: a) REF AK-04220, UDI/DI 20801902190943, Lot Numbers: 13F22E0436, 13F22H0360, 33F23E0375, 33F23K0032, 33F23J0443, 33F23L0862; UDI/DI 30801902121616, Lot Numbers: 13F22L0105, 13F23A0651, 33F23C0100, 33F23F0347, 33F23E0501, 33F23G0141; b) REF ASK-04220-FMH1, UDI/DI 20801902111665, Lot Numbers: 33F23B0525, 33F23H0100; c) REF ASK-04220-HHC, UDI/DI 20801902190851, Lot Numbers: 33F23F0067, 33F23K0278, 33F23G0478, 33F23H0101; d) REF ASK-04220-KSP, UDI/DI 20801902156406, Lot Numbers: 33F23E0709, 33F23H0102, 33F23H0918, 33F23J0607; e) REF ASK-04220-PSU, UDI/DI 40801902121446, Lot Numbers: 33F23F0331, 33F23G0663; f) REF ASK-04220-UCL1, UDI/DI 20801902159124, Lot Numbers: 33F23F0801, 33F23C0726, 33F23H0320; g) REF ASK-04500-AH, UDI/DI 20801902114048, Lot Numbers: 33F23E0731, 33F23H0103, 33F23H0395; h) REF ASK-04500-HF-S, UDI/DI 40801902118002, Lot Numbers: 33F23F0202, 33F23H0104, 33F23H0396; i) REF NA-04220-S1A, UDI/DI 20801902139416, Lot Numbers: 33F23J0671; j) REF NA-04220-X1A, UDI/DI 20801902136606, Lot Numbers: 33F23F0704; k) REF RA-04220, UDI/DI 20801902180678, Lot Numbers: 14F23B0041; UDI/DI 30801902193309, Lot Numbers: 14F21L0189, 14F21L0190, 14F21L0250, 14F21L0251, 14F21L0252, 14F21L0253, 14F21M0130, 14F21M0131, 14F21M0205, 14F21M0206, 14F22A0044, 14F22A0123, 14F22A0124, 14F22A0143, 14F22A0144, 14F22B0163, 14F22B0164, 14F22B0175, 14F22C0017, 14F22C0018, 14F22C0019, 14F22C0020, 14F22C0021, 14F22C0022, 14F22C0216, 14F22C0284, 14F22C0346, 14F22C0347, 14F22D0023, 14F22D0024, 14F22D01

UPC codes

20801902190943, 30801902121616, 20801902111665, 20801902190851, 20801902156406, 40801902121446, 20801902159124, 20801902114048, 40801902118002, 20801902139416, 20801902136606, 20801902180678, 30801902193309, 30801902210631, 30801902193316, 30801902202988

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify