FDA (Device enforcement)Recalled 2024-01-29class ii

Canon Medical System, USA, INC.: MRI system: Vantage Elan, Model: MRT-2020

Hazard

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

Details from the source

Reason for recall: For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan. Distribution: Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO. Quantity: 1 Lot/code info: UDI-DI: 04987670101632. Serial Number: S1F2052753

UPC codes

04987670101632

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify