← Search recalls
FDA (Device enforcement)Recalled 2023-02-01class ii

Scientia Vascular, Inc.: Plato 17 Microcatheter: Straight Tip, PL17-160-000 ; 45 degree Tip, PL17-160-045; 90 degree Tip, PL17-160-090

Hazard

Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.

Details from the source

Reason for recall: Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response. Distribution: US Nationwide distribution in the states of IL, AL, FL, NC, WI, GA, SC, IN, KS, CA, OH, NY, NH, CO, MA, NV, UT, TX. Quantity: 353 Lot/code info: All lots, REF/UDI-DI/Lot(Expiration): PL17-160-000/00818075010455/ 900164(8/31/2022), 900165(8/31/2022), 900172(8/31/2022), 900177(9/30/2022), 900180(9/30/2022), 900186(10/31/2022), 900187(10/31/2022), 900188(10/31/2022), 900193(10/31/2022), 900196(12/31/2022), 900199(12/31/2022), 900202(12/31/2022), 900221(2/28/2023), 900222(2/28/2023), 900225(2/28/2023), 900226(3/31/2023), 20441(11/30/2023), 20476(11/30/2023), 20650(12/31/2023); PL17-160-045/00818075010462/900167(8/31/2022), 900168(8/31/2022), 900169(8/31/2022), 900174(8/31/2022), 900175(8/31/2022), 900179(9/30/2022), 900181(9/30/2022), 900183(10/31/2022), 900185(10/31/2022), 900205(1/31/2023), 900206(1/31/2023), 900210(1/31/2023), 900211(1/31/2023), 900215(1/31/2023), 900220(1/31/2023), 900219(1/31/2023), 900217(2/28/2023), 900223(2/28/2023), 900224(2/28/2023); PL17-160-090/00818075010479/900170(8/31/2022), 900171(8/31/2022), 900178(9/30/2022), 900182(9/30/2022), 900184(10/31/2022), 900192(11/30/2022), 900195(11/30/2022), 900209(1/31/2023), 900228(3/31/2023)

UPC codes

00818075010455, 00818075010462, 00818075010479

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.