There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Reason for recall: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection. Distribution: Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam. Quantity: 14,789 (8651 US; 6138 OUS) Lot/code info: 1. Flat Vitrectomy Lens Pouch UDI 810123482290 Box UDI 810123482306 Lots 2410036 2411020 2. Volk Magnifying Lens Pouch UDI 810123482825 Box UDI 810123482832 Lot 2411019
810123482290, 810123482306, 810123482825, 810123482832
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.