Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
Reason for recall: Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted. Distribution: Nationwide US distribution. Worldwide international distribution. Quantity: 11,792 units (5.408 US, 6,384 OUS) Lot/code info: Model Number (UDI-DI): D998-00-0800-31 (10607567109053¿), D998-00-0800-32 (10607567111117), D998-00-0800-33 (10607567109008), D998-00-0800-35 (10607567109107), D998-00-0800-36 (10607567114187), D998-00-0800-45 (10607567108421), D998-00-0800-52 (10607567108438), D998-00-0800-53 (10607567108391), D998-00-0800-55 (10607567108414), D998-00-0800-65 (10607567113432), D998-00-0800-83 (10607567108407), D998-UC-0800-31 (10607567109053), D998-UC-0800-33 (10607567109008), D998-UC-0800-52 (10607567108438), D998-UC-0800-53 (10607567108391), D998-UC-0800-55 (10607567108414). All serial numbers
10607567109053, 10607567111117, 10607567109008, 10607567109107, 10607567114187, 10607567108421, 10607567108438, 10607567108391, 10607567108414, 10607567113432, 10607567108407