Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.
Reason for recall: Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death. Distribution: Worldwide distribution - US Nationwide distribution in the states of Arkansas, California, Florida, Georgia, Louisiana, Massachusetts, New York, North Carolina, Ohio, Washington and the countries of Australia, Austria, Belgium, Canada, Canary Islands, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, New Caledonia, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Quantity: 264 US; 5056 OUS Lot/code info: UDI-DI (GTIN): 20884521543598 Lots: N2D0002Y, N2D0004Y, N2D0195Y