Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reorder Pack, REF: 250-3000
Check against the official notice below — it lists the full affected range.
Bio-Rad Laboratories is recalling specific lots of the VARIANT nbs Sickle Cell Program Reorder Pack containing a Resin Update CD-ROM (Model 250-3020) due to a software problem. When the update procedure is performed, the CD-ROM overwrites customized settings in the Setup/Test/Data Setup and Pattern Setting subscreens with default settings. This creates a risk that some patient samples could be assigned different sickle cell patterns than what the laboratory has validated using its custom settings. Bio-Rad received complaints from four customers who experienced this issue before initiating the recall.
The affected product was distributed to laboratories across the United States, Canada, Spain, France, Italy, India, Vietnam, and the Philippines. Approximately 1,893 units are involved, identified by specific lot numbers for both the CD-ROM (lots AX90085, AY90085, AZ90085, BB90085, CB90085, DB90085) and the reorder packs with various lot numbers and expiration dates through May 2023. The recall record does not specify a remedy or action that affected consumers should take.
Written by software from the official notice below. Not reviewed by a lawyer.
The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in specific lots of VARIANT nbs Sickle Cell Progam Reorder Pack causes all customized settings in the Setup/Test/Data Setup subscreen and Setup/Test/Pattern Setting subscreen to be overwritten with the default settings when the Update Kit procedure is performed.The VARIANTnbs Sickle Cell Program has been validated using the default Resin Update CD¿ROM parameters for the intended use. However, there is a risk that the customer's custom pattern settings may be different than the default settings, resulting in a different pattern being assigned to some samples that the laboratory has validated. Following the complaints from 4 customers who experienced these issues, the firm initiated this recall.
Reason for recall: The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in specific lots of VARIANT nbs Sickle Cell Progam Reorder Pack causes all customized settings in the Setup/Test/Data Setup subscreen and Setup/Test/Pattern Setting subscreen to be overwritten with the default settings when the Update Kit procedure is performed.The VARIANTnbs Sickle Cell Program has been validated using the default Resin Update CD¿ROM parameters for the intended use. However, there is a risk that the customer's custom pattern settings may be different than the default settings, resulting in a different pattern being assigned to some samples that the laboratory has validated. Following the complaints from 4 customers who experienced these issues, the firm initiated this recall. Distribution: US: AZ, AR, GA, CO, TN, LA, VA, FL, KY, IN, KS, MD, MN, MO, NC, NJ, NY, OH, OR, SC, IA, AL, UT, PR OUS: Canada, Spain, France, Italy, India, Vietnam, Philippines Quantity: 1893 Lot/code info: REF: 250-3020/Lot: AX90085, AY90085, AZ90085, BB90085, CB90085, DB90085; REF: 250-3000, UDI/DI: 00847817008276, Lot/Expiry: 64442741/ 2/28/2023, 64444208/ 2/28/2023, 64444218/ 3/26/2023, 64444715/ 3/31/2023, 64452929/ 3/31/2023, 64452922/ 5/31/2023