← Search recalls
FDA (Device enforcement)Recalled 2025-12-26class ii

Cardinal Health 200, LLC: Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 3.

Hazard

The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.

Details from the source

Reason for recall: The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization. Distribution: Worldwide distribution - US Nationwide and the countries of Brazil, Canada, Chile, Colombia, Dominican Republic, and Mexico. Quantity: 1,169,726 units Lot/code info: All units affected by IFU update. UDI-DI: 1. 8888571562 Cardinal Health Sentinel Seal CDU 50192253003091 (CS) 10192253003093 (EA) 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 50192253003084 (CS) 10192253003086 (EA) 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 50192253003077 (CS) 10192253003079 (EA) 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 50192253003039 (CS) 10192253003031 (EA) 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 50192253003015 (CS) 10192253003017 (EA) 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 50192253003060 (CS) 10192253003062 (EA) 7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy Change Connector 50192253003022 (CS) 10192253003024 (EA) 8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL 50192253002643 (CS) 10192253002645 (EA) 9. 1814713105 OUS¿Only - Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL (Bottle only) 50192253002308 (CS) 10192253002300 (CS) 10. 1180571570 OUS¿Only - Cardinal Health Thora-Seal CDU, Two Chamber System, 2600 mL 50192253002292 (CS) 10192253002294 (EA) 11. 8884713308 Cardinal Health Thora-Seal CDU, Three Chamber System, 2500 mL 50192253002650 (CS) 10192253002652 (EA) 12. 8884713900 Cardinal Health Thora-Seal Replacement Collection Chamber, 2500 mL 50192253002667 (CS) 10192253002669 (EA)

UPC codes

50192253003091, 10192253003093, 50192253003084, 10192253003086, 50192253003077, 10192253003079, 50192253003039, 10192253003031, 50192253003015, 10192253003017, 50192253003060, 10192253003062, 50192253003022, 10192253003024, 50192253002643, 10192253002645, 50192253002308, 10192253002300, 50192253002292, 10192253002294, 50192253002650, 10192253002652, 50192253002667, 10192253002669

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.