There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
Reason for recall: There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU. Distribution: US Nationwide distribution in the states of TX, AL, NV; AU, NZ. Quantity: 3 units Lot/code info: UDI-DI UDI-DI 00855913001848, Serial Numbers: 1304384, 1304619, 1304620
00855913001848