Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
Reason for recall: Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs. Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Belgium ENG, Canada, China, China APactron, Denmark, France, Germany, Hong Kong S.A.R. China, India, Italy, Japan, Korea, Netherlands, Norway, Poland, Singapore, Spain, SwitzerlandENG, Taiwan, Thailand, UK. Quantity: 6 Lot/code info: Lot Code: UDI: 0735000201038920210518, 0735000201043320210610, 0735000201044020210916 and 0735000201063120220616 GTIN: 07350002010389, 07350002010433, 07350002010440 and 07350002010631 Serial Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116 and 11.0.4.15 Software Revisions: RayStation 11A and 11A Service Pack 1, 2 and 3 Expiration Date: 2027-06-22
07350002010389, 07350002010433, 07350002010440, 07350002010631
Official record: FDA (Device enforcement) ↗