FDA (Device enforcement)Recalled 2024-02-06class ii

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Azurion 3 with a Certeray generator -To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722221 (2) 7

Hazard

Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.

Details from the source

Reason for recall: Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure. Distribution: Worldwide distribution - US Nationwide distribution in the states of NY, PA, PR, TN, TX, VA, WA and the countries of Algeria, Armenia, Austria, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Macao, Macedonia, Malaysia, Morocco, Netherlands, Norway, Oman, Palestine, Quantity: 14 units Lot/code info: OUS: (1) Model: 722221 SN: 66 (01)00884838099203(21)66; SN: 91 (01)00884838099203(21)91; SN: 104 (01)00884838099258(21)104, (2) Model: 722222 SN: UDI: 165 (01)00884838099210(21)165 183 (01)00884838099210(21)183 189 (01)00884838099210(21)189 182 (01)00884838099210(21)182 185 (01)00884838099210(21)185 159 (01)00884838099210(21)159 184 (01)00884838099210(21)184 190 (01)00884838099210(21)190 174 (01)00884838099210(21)174 177 (01)00884838099210(21)177 (3)Model: 722280 20 (01)00884838103276(21)20

UPC codes

00884838099203, 00884838099258, 00884838099210, 00884838103276

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify