FDA (Device enforcement)Recalled 2022-12-21class ii

Siemens Medical Solutions USA, Inc: Luminos dRF Max with software VE10, VF10, VF11- A diagnostic imaging system for radiographic and fluoroscopic studie Model Number:10762471

Hazard

Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values

Details from the source

Reason for recall: Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values Distribution: US Nationwide distribution. Quantity: 2 units; Expanded recall: 53 units Lot/code info: UDI: 4056869009162 S/N: 5037 5951 Expanded Recall 2/24/23: 5069 5085 5209 5234 5340 5364 5634 5664 5682 5919 5920 5921 5947 5953 5954 5956 5963 7075 7098 7147 7162 7168 7172 7208 7266 7310 7319 7391 7404 7442 7460 7488 7545 7743 7771 8074 8093 8094 8183 8198 8223 8242 8244 8275 8320 8322 8334 8361 8371 8378 8385 8389

UPC codes

4056869009162

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.