FDA (Device enforcement)Recalled 2024-01-04class ii

MEDLINE INDUSTRIES, LP - Northfield: Medline Contro-Bulb Syringe, REF DYND20125; bulb irrigation syringe

Hazard

Packaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during transportation. This only affects two lot numbers (96922110001 and 96923100001).

Details from the source

Reason for recall: Packaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during transportation. This only affects two lot numbers (96922110001 and 96923100001). Distribution: US, Nationwide Quantity: 4831 units Lot/code info: UDI/DI 40080196031765 (case), 10080196031764 (unit) Lot Numbers: 96922110001 and 96923100001

UPC codes

40080196031765, 10080196031764, 96922110001, 96923100001

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.