FDA (Device enforcement)Recalled 2020-01-06closedclass ii

MEDTECH SAS: ROSA One 3.1 Brain Application

Is my product affected?

ROSA One 3.1 Brain Application

  • UPC 03760244030948
  • UPC 03760244032492

Check against the official notice below — it lists the full affected range.

What happened

MedTech SAS has recalled the ROSA One 3.1 Brain Application due to a software issue affecting image reconstruction and display. When using ROSA Brain software, some cross-sectional images from patient head scans may not be reconstructed or displayed properly in two and three-dimensional views, which could compromise surgical planning. The recall affects 39 units distributed nationwide across 12 specific serial numbers.

The record does not specify a remedy or action that affected consumers should take.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.

Details from the source

Reason for recall: Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning. Distribution: US Nationwide Quantity: 39 units Lot/code info: Serial Number / UDI Number: 1) BS19037 / UDI - (01)03760244030948(11)190423(21)BS19037; 2) BS19038 / UDI - (01)03760244030948(11)190415(21)BS19038; 3) BS19039 / UDI - (01)03760244030948(11)190430(21)BS19039; 4) BS19040 / UDI - (01)03760244030948(11)190513(21)BS19040; 5) BS19041 / UDI - (01)03760244030948(11)190523(21)BS19041; 6) BS19043 / UDI - (01)03760244030948(11)190513(21)BS19043; 7) BS19049 / UDI - (01)03760244032492(11)190730(21)BS19049; 8) BS19050 / UDI - (01)03760244032492(11)190726(21)BS19050; 9) BS19052 / UDI - (01)03760244032492(11)190731(21)BS19052; 10) BS19054 / UDI - (01)03760244032492(11)190912(21)BS19054; 11) BS19057 / UDI - (01)03760244032492(11)191003(21)BS19057; 12) BS19058 / UDI - (01)03760244032492(11)191011(21)BS19058.

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify