Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
Reason for recall: Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples. Distribution: Nationwide Foreign: Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Bulgaria¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Cura¿ao,¿St¿Eus Cyprus¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hungary¿¿¿¿¿¿¿¿ Iran¿¿¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Jordan¿¿¿¿¿¿¿¿¿ Kenya¿¿¿¿¿¿¿ Quantity: 13,843 kits Lot/code info: (01)00630414598963(10)15308337(17)20230320 (01)00630414598963(10)27949339(17)20230529 (01)00630414598963(10)41143343(17)20230630 (01)00630414598963(10)41144343(17)20230630 (01)00630414598963(10)62969345(17)20230726 Lot Numbers: 15308337, 27949339, 41143343, 41144343, 62969345 and higher. Affects all current and future lots of the Atellica IM Folate assay until the Instructions for Use are updated.