← Search recalls
FDA (Device enforcement)Recalled 2024-01-25class ii

Voco GmbH: IonoStar Plus, Catalog (REF) Numbers: a) 2540, b)2543, c) 2544, d) 2545, e) 2546, f) 2547, g) 2549; dental cement

Hazard

Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.

Details from the source

Reason for recall: Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible. Distribution: Worldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS (foreign) to countries of: Belgium, Bulgaria, Denmark, Germany, France, Greece, Great Britain, India, Italy, Canada, Quatar, Lithuania, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic, Australia, Saudi Arabia, & Ireland. Quantity: 1530 units Lot/code info: a) 2540, UDI/DI +E22125401, Lot Numbers: 2338699, 2347154; b)2543, UDI/DI +E22125431, Lot Numbers: 2344046; c) 2544, UDI/DI +E22125441, Lot Numbers: 2342759, 2439537; d) 2545, UDI/DI +E22125451, Lot Numbers: 2348503; e) 2546, UDI/DI +E22125461, Lot Numbers: 2344045, 2349452; f) 2547, UDI/DI +E22125471, Lot Numbers: 2349453; g) 2549, UDI/DI +E22125491, Lot Numbers: 2342760, 2344016, 2344047, 2351233

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify