← Search recalls
FDA (Device enforcement)Recalled 2020-01-29closedclass i

Abbott Vascular: NC TRAVELER Coronary Dilatation Catheter, REF 1013157-15, Rx Only,NC Traveler RX 4.0 X 15MM The NC Traveler RX Coronary Dilatation Catheter is not cleared for commercial distribution

Is my product affected?

NC TRAVELER Coronary Dilatation Catheter, REF 1013157-15, Rx Only,NC Traveler RX 4.0 X 15MM The NC Traveler RX Coronary Dilatation Catheter is not cleared for commercial distribution in the United States.

  • UPC 08717648196003

Check against the official notice below — it lists the full affected range.

What happened

Abbott Vascular is recalling the NC Traveler RX 4.0 X 15MM Coronary Dilatation Catheter (Part Number 1013157-15, Lot Number 91102G1) because the device may have difficulty or inability to deflate the balloon. A total of 791 units were distributed worldwide, with all affected units distributed outside the United States to countries including Algeria, Argentina, Australia, Austria, Brazil, Canada, China, Colombia, and many others.

The FDA record does not specify a remedy or instructions for affected consumers regarding this recall.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.

Details from the source

Reason for recall: Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon. Distribution: Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federa Quantity: Total 791 Traveler (0 US and 791 OUS) Lot/code info: Device Identifier/GTIN: 08717648196003 Part Number: 1013157-15 Lot Numbers:91102G1

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify