Due to a trend in false positive Candida tropicalis results on blood culture panel when using certain BD BACTEC vial mediums
Reason for recall: Due to a trend in false positive Candida tropicalis results on blood culture panel when using certain BD BACTEC vial mediums Distribution: Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Andorra, United Arab Emirates, Armenia, Argentina, Austria, Australia, Bosnia, Belgium, Quantity: 32,910 panels/pouches Lot/code info: For BioFire Blood Culture Identification 2 (BDID2) Panel Part Number: RFIT-ASY-0147 UDI-DI code: 00815381020338 Lot Numbers: All non-expired BCID2 Panel lots, if used with the BD BACTEC vials including but not limited to, the bottle types listed in Table 1. NOTE: The affected blood culture vial types most likely contain non-viable Candida tropicalis nucleic acid/DNA fragments that can be detected by the BCID2 Panel. Therefore, all lots of the BioFire BCID2 Panel would be affected if they are used with the affected blood culture vials. Neither the BioFire BCID2 test kits nor the BD BACTEC vials have a defect - both products are performing as intended.
00815381020338