Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).
Reason for recall: Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL). Distribution: US: MD, NJ, OH, FL, CA, TX, NC, WA, MA, NV, GA, CO, PA, NM, UT, VA, KS, MT, OK, NY, AZ, LA, OR, AL, MO, IN, IL, KY, MS, NH, AK, SD, MN, TN, PR, MI, IA, WI, WV, ME, SC, HI, RI, ND, NE, ID, AR, WY, DC, GU, DE, CT. OUS: Brazil, South Africa, Argentina, Mexico, Canada, Costa Rica, Peru, El Salvador, Ecuador, Panama, Singapore, Ireland, United Kingdom of Great Britain and Northern Ireland, Germany, Chi Quantity: 34,451 Lot/code info: REF/UDI-DI: OSR6196/15099590010928, OSR6296/15099590020361. All Lots and Expiration Dates. IFU Version BAOSR6X96-10 and prior.
15099590010928, 15099590020361
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.