Sysmex America, Inc.: Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing works

Hazard

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

Details from the source

Reason for recall: Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System. Distribution: US Nationwide. Quantity: 32 units Lot/code info: UDI-DI: 04987562501908; Serial Numbers: 1908002275 1908002276 1908009775 1910009588 1910011295 1912002638 2103013913 2108004629 2108004630 2108004631 2108004632 2112005423 2112005425 2203014206 2203014207 2203014208 2205012732 2205012734 2206005973 2206012294 2208015835 2210003722 2210003723 2210003724 2212010048 1903011966 1903011968 1910009589 1912002640 2103013912 2208015834 1903011967

UPC codes

04987562501908

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.