Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Reason for recall: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. Distribution: Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS Quantity: 0 Lot/code info: Lot Code: UDI: 0735000201047120220131 GTIN: 07350002010471 Serial Numbers: 10.1.1.54 Software Revision: RayStation 10B SP1 Expiration Date: 2027-02-01