FDA (Device enforcement)Recalled 2025-11-28class ii

RAYSEARCH LABORATORIES AB: RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B S

Hazard

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Details from the source

Reason for recall: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. Distribution: Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS Quantity: 22 units Lot/code info: UDI: 0735000201042620211208, 0735000201049520220312, 0735000201050120220422, 0735000201060020220620 and 0735000201057020221222. GTIN: 07350002010426, 07350002010495, 07350002010501, 07350002010600 and 07350002010570. Serial Numbers: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12, 12.3.0.119. Software Revisions: RayStation 11B, 11B SP1, 11B SP2, 11B SP3, 11B SPT1. Expiration Date: 2028-03-20.

UPC codes

07350002010426, 07350002010495, 07350002010501, 07350002010600, 07350002010570

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify