The MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD).
Reason for recall: The MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD). Distribution: Worldwide distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Germany, Italy, Netherlands, Portugal, Switzerland, and United Kingdom. Quantity: 23 units (6 units US) Lot/code info: Model numbers MSOT Acuity Echo (Research System) MSOT Acuity Echo CE MSOT Acuity Echo CE2 UDI's: (01)04262380070004 (01)04262380070004(21)22102 (01)04262380070004(21)22203 (01)04262380070004(21)21901 (01)04262380070004(21)22101 (01)04262380070004(21)22104 (01)04262380070004(21)22105 (01)04262380070004(21)22106 (01)04262380070004(21)22107 (01)04262380070004(21)22201 (01)04262380070004(21)22202 (01)04262380070004(21)22204 (01)04262380070004(21)22205 (01)04262380070004(21)22206 (01)04262380070004(21)22301 (01)04262380070004(21)22401