FDA (Device enforcement)Recalled 2025-10-10class ii

Thoratec LLC: CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 20

Hazard

Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.

Details from the source

Reason for recall: Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death. Distribution: Worldwide - US Nationwide distribution including in the states of MN, PA, OK, FL, CA, OR, IL, NY, DE, GA, NJ, NC, WI, OH, AZ, AR, TX, TN, MO, MA, MT, KY, NE, WV, VA, ME, ND, AL, CO, DC, MS, IN, MI, CT, SC, WA, NM, MD, HI, ID, IA, LA, UT, RI, NV and the countries of Algeria, Australia, Austria, Bahrain, CANADA, Cayman Islands, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, F Quantity: 8399 Lot/code info: UDI-DI: 07640135140627, 07640135140603, 05415067036414 IFU: Part(Name) Revision: For US: 1. ARTEN600254920 (IFU CentriMag Blood Pump ABT) Rev C 2. ARTEN600254958 (IFU PediMag Blood Pump US ABT) Rev C 3. ARTEN100174794 (CentriMag Adult Pre-connected Pack IFU, US) Rev B 4. ARTEN600230157 (IFU CentriMag Blood Pump ECMO US) Rev D FOR OUS: ARTEN600254935 (IFU CentriMag Blood Pump CAN ABT) Rev C ARTEN600254936 (IFU PediVas Blood Pump Canada ABT) Rev C ARTEN600332205 (IFU CMag VAS-ECMO Pump OUS) Rev A ARTEN600332207 (IFU PVAS VAS-ECMO Pump OUS) Rev A REF 201-20003, 104318, 201-51006, 201-51007, 201-90001 201-90001T don't have associated un-expired batches/serial numbers. REF/Batch(Serial): 102953/10007334, 10007335, 10007336, 10012012, 10012013, 10012014, 10012015, 10012016, 10036179, 10036180, 10036181, 10037990, 10037992, 10037993, 10037995, 10037996, 10046028, 10046029, 10046030, 10046031, 10046032, 10046033, 10058754, 10058755, 10058756, 10061673, 10061675, 10061676, 10061677, 10065946, 10065947, 10065954, 10065955, 10065956, 10071358, 10071359, 10071361, 10073780, 10073781, 10073783, 10082053, 10082055, 10103208, 10103209, 10106345, 10106346, 10106347, 10106348, 10120074, 10120075, 10120076, 10120077, 10122157, 10122158, 10122159, 10125562, 10125563, 10125564, 10125565, 10138336, 10138337, 10138338, 10138339, 10145124, 10145125, 10145126, 10145127, 10164040, 10164041, 10164044, 10168193, 10168194, 10168195, 10168196, 10168197, 10168199, 10168200, 10168201, 10168202, 10168203, 1

UPC codes

07640135140627, 07640135140603, 05415067036414

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify