Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.
Reason for recall: Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration. Distribution: Worldwide - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, Germany, Jordan, Lebanon, Malaysia, Netherlands, New Zealand, Peru, Poland, Russia, Singapore, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam. Quantity: 442 Lot/code info: UDI-DI (GTIN): 00763000547332 00763000547271 00763000547318 00763000547264 00643169796225 00643169796300 00643169796294 00643169796232 00643169796287 00763000547325 00643169796300 00763000547257 Serial/Lot numbers C112511 C113253 C113795 C117416 C117482 C121084 C123953 C133327 C133329 C134478 C138060 C150325
00763000547332, 00763000547271, 00763000547318, 00763000547264, 00643169796225, 00643169796300, 00643169796294, 00643169796232, 00643169796287, 00763000547325, 00763000547257