FDA (Device enforcement)Recalled 2025-12-12class ii

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model

Hazard

Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.

Details from the source

Reason for recall: Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error. Distribution: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican R Quantity: 10,466 units Lot/code info: Software Versions: R2.1.10 and R2.2.10. 1. Model Number (UDI): 722063 ((01)00884838085275(21)122, (01)00884838085275(21)64, (01)00884838085275(21)56, (01)00884838085275(21)131, (01)00884838085275(21)84, (01)00884838085275(21)132, (01)00884838085275(21)96, (01)00884838085275(21)98, (01)00884838085275(21)108, (01)00884838085275(21)153, (01)00884838085275(21)69, (01)00884838085275(21)154, (01)00884838085275(21)97, (01)00884838085275(21)137, (01)00884838085275(21)116, (01)00884838085275(21)127, (01)00884838085275(21)151, (01)00884838085275(21)148, (01)00884838085275(21)150, (01)00884838085275(21)146, (01)00884838085275(21)149, (01)00884838085275(21)123, (01)00884838085275(21)130, (01)00884838085275(21)143, (01)00884838085275(21)139, (01)00884838085275(21)144, (01)00884838085275(21)145, (01)00884838085275(21)141, (01)00884838085275(21)140, (01)00884838085275(21)135, (01)00884838085275(21)133, (01)00884838085275(21)107, (01)00884838085275(21)120, (01)00884838085275(21)129, (01)00884838085275(21)105, (01)00884838085275(21)110, (01)00884838085275(21)128, (01)00884838085275(21)119, (01)00884838085275(21)113, (01)00884838085275(21)43, (01)00884838085275(21)117, (01)00884838085275(21)115, (01)00884838085275(21)101, (01)00884838085275(21)94, (01)00884838085275(21)106, (01)00884838085275(21)103, (01)00884838085275(21)99, (01)00884838085275(21)102, (01)00884838085275(21)104, (01)00884838085275(21)100, (01)00884838085275(21)80, (01)00884838085275(21)81, (01)00884838085275(21)89, (01)0088483

UPC codes

00884838085275, 00884838099203, 00884838085282, 00884838099210, 00884838103276, 00884838099227

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.