← Search recalls
FDA (Device enforcement)Recalled 2025-12-16class ii

Medline Industries, LP: Medline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF CDS980647K; 3) DAVINCI PROCEDURE, REF CDS980864Q; 4) GYN-URO LAPSCP CDS-L

Hazard

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Details from the source

Reason for recall: Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use. Distribution: US Nationwide distribution. Quantity: 12584 units Lot/code info: REF CDS980523M: UDI/DI 10195327508913 (EA) 40195327508914 (CS), Lot number 25FBF036; REF CDS980523M: UDI/DI 10195327508913 (EA) 40195327508914 (CS), Lot number 25DBB218; REF CDS980647K: UDI/DI 10195327508920 (EA) 40195327508921 (CS), Lot number 25DBP829; REF CDS980647K: UDI/DI 10195327508920 (EA) 40195327508921 (CS), Lot number 25CBI860; REF CDS980864Q: UDI/DI 10198459121937 (EA) 40198459121938 (CS), Lot number 25EMC533; REF CDS981583U: UDI/DI 10195327209766 (EA) 40195327209767 (CS), Lot number 25CBC570; REF CDS981583V: UDI/DI 10198459311390 (EA) 40198459311391 (CS), Lot number 25FBK816; REF CDS981583V: UDI/DI 10198459311390 (EA) 40198459311391 (CS), Lot number 25EBS462; REF CDS981826K: UDI/DI 10198459224805 (EA) 40198459224806 (CS), Lot number 25FBH633; REF CDS981826K: UDI/DI 10198459224805 (EA) 40198459224806 (CS), Lot number 25BBU420; REF CDS982377K: UDI/DI 10198459229039 (EA) 40198459229030 (CS), Lot number 25DMI839; REF CDS982377K: UDI/DI 10198459229039 (EA) 40198459229030 (CS), Lot number 25DMF861; REF CDS982377K: UDI/DI 10198459229039 (EA) 40198459229030 (CS), Lot number 25CMC823; REF CDS982779N: UDI/DI 10195327647711 (EA) 40195327647712 (CS), Lot number 25EBF365; REF CDS982779N: UDI/DI 10195327647711 (EA) 40195327647712 (CS), Lot number 25CBM352; REF CDS982780V: UDI/DI 10195327647728 (EA) 40195327647729 (CS), Lot number 25EBI642; REF CDS982780V: UDI/DI 10195327647728 (EA) 40195327647729 (CS), Lot number 25DBP916; REF CDS982780V: UDI/DI 10195327647728 (EA) 401953276477

UPC codes

10195327508913, 40195327508914, 10195327508920, 40195327508921, 10198459121937, 40198459121938, 10195327209766, 40195327209767, 10198459311390, 40198459311391, 10198459224805, 40198459224806, 10198459229039, 40198459229030, 10195327647711, 40195327647712, 10195327647728, 40195327647729, 10193489927863, 40193489927864, 10198459224614, 40198459224615, 10198459107368, 40198459107369, 10198459312113, 40198459312114, 10198459019418, 40198459019419, 10195327689124, 40195327689125, 10195327677145, 40195327677146, 10195327543693, 40195327543694, 10198459331923, 40198459331924, 10195327500993, 40195327500994, 10195327261504, 40195327261505

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.