← Search recalls
FDA (Device enforcement)Recalled 2025-12-16class ii

Medline Industries, LP: Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) T

Hazard

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Details from the source

Reason for recall: Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use. Distribution: US Nationwide distribution. Quantity: 4536 units Lot/code info: REF CDS860237AF: UDI/DI 10198459232626 (EA) 40198459232627 (CS), Lot Number 25FBK870; REF CDS860237AF: UDI/DI 10198459232626 (EA) 40198459232627 (CS), Lot Number 25EBD110; REF CDS982128Q: UDI/DI 10198459008245 (EA) 40198459008246 (CS), Lot Number 25EBC253; REF CDS982128Q: UDI/DI 10198459008245 (EA) 40198459008246 (CS), Lot Number 25DBG393; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25FBQ303; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25EBD631; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25DBM322; REF CDS984640I: UDI/DI 10198459318955 (EA) 40198459318956 (CS), Lot Number 25DBT807; REF DYNJ0525321O: UDI/DI 10198459236907 (EA) 40198459236908 (CS), Lot Number 25DMG099; REF DYNJ0525321O: UDI/DI 10198459236907 (EA) 40198459236908 (CS), Lot Number 25CMJ006; REF DYNJ57640F: UDI/DI 10198459011115 (EA) 40198459011116 (CS), Lot Number 25FDB347; REF DYNJ62594D: UDI/DI 10198459049767 (EA) 40198459049768 (CS), Lot Number 25EBK539; REF DYNJ62594D: UDI/DI 10198459049767 (EA) 40198459049768 (CS), Lot Number 25DBD145; REF DYNJ63392D: UDI/DI 10198459164507 (EA) 40198459164508 (CS), Lot Number 25BLA967; REF DYNJ68187B: UDI/DI 10198459382086 (EA) 40198459382087 (CS), Lot Number 25FMG440; REF DYNJ82368A: UDI/DI 10198459258077 (EA) 40198459258078 (CS), Lot Number 25EMH020; REF DYNJ82368A: UDI/DI 10198459258077 (EA) 40198459258078 (CS), Lot Number 25CMF917; REF DYNJ85261: UDI/DI 10195327483418 (EA) 4019532

UPC codes

10198459232626, 40198459232627, 10198459008245, 40198459008246, 10198459148101, 40198459148102, 10198459318955, 40198459318956, 10198459236907, 40198459236908, 10198459011115, 40198459011116, 10198459049767, 40198459049768, 10198459164507, 40198459164508, 10198459382086, 40198459382087, 10198459258077, 40198459258078, 10195327483418, 40195327483419, 10198459104800, 40198459104801, 10198459047817, 40198459047818, 10198459157936, 40198459157937, 10193489851588, 40193489851589, 10198459037863, 40198459037864, 10198459300592, 40198459300593, 10198459332883, 40198459332884, 10888277776760, 40888277776761, 10193489569032, 40193489569033

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.