FDA (Device enforcement)Recalled 2022-02-11class ii

PHILIPS MEDICAL SYSTEMS: CombiDiagnost R90 1.0 (709030)

Is my product affected?

CombiDiagnost R90 1.0 (709030)

  • UPC 00884838076747

Check against the official notice below — it lists the full affected range.

What happened

Philips Medical Systems is recalling 86 units of the CombiDiagnost R90 1.0 fluoroscopy imaging system due to a software issue where the Radio Fluoroscopy viewer may display a previous patient's radiography images during an examination. When this occurs, the displayed images differ in content, format, and size from the current examination images. The affected devices are distributed nationwide across 16 states and are identified by specific serial numbers with software version 1.1.1 or higher.

The recall notice does not specify a remedy for affected consumers or healthcare facilities.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.

Details from the source

Reason for recall: While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size. Distribution: US Nationwide distribution in the states of CA, FL, GA, IL, IN, KS, MD, ME, MI, MN, NC, NJ, NY, TX, VA, WA. Quantity: 86 units Lot/code info: Model: R90 1.0 (709030); UDI DI: 00884838076747; Software version 1.1.1 or higher; Serial numbers: 10000000, 10000002, 10000003, 10000004, 10000006, 10000011, 10000012, 10000020, 10000024, 10000027, 10000049, 10000050, 10000065, 10000066, 10000067, 10000069, 10000078, 10000081, 10000099, 10000103, 10000107, 10000140, 10000145, 10000152, 10000180, 10001001, 10001010, 10001095, 591362, SN16000009, SN17000003, SN17000007, SN17000015, SN17000022, SN17000111, SN17000114, SN17000118, SN17000120, SN17000121, SN17000122, SN17000123, SN17000128, SN17000129, SN17000130, SN17000145, SN17000146, SN17000147, SN17000148, SN17000153, SN17000157, SN17000163, SN17000165, SN17000166, SN18000003, SN18000004, SN18000010, SN18000021, SN18000025, SN18000033, SN18000037, SN18000038, SN18000044, SN18000045, SN18000053, SN18000054, SN18000057, SN18000060, SN18000068, SN18000069, SN18000079, SN19000004, SN19000006, SN19000011, SN19000020, SN19000034, SN19000036, 27, 40, 41, 42, 77, 101, 131, 155, 180, 190;

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.