MicroScan WalkAway-40 plus Reconditioned Instrument REF B1018-340R
Check against the official notice below — it lists the full affected range.
Beckman Coulter Inc. is recalling three MicroScan WalkAway-40 plus Reconditioned Instruments (catalog number B1018-340R) that may have been assembled with an incorrect AC power entry connector inlet. The connector is rated for 1 ampere of current instead of the correct 10 ampere rating. The affected instruments have serial numbers 34020035, 34020224, and 34030054. These units were distributed worldwide and within the U.S. in Alaska, Arkansas, California, Colorado, the District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington.
The recall notice does not specify a remedy or instructions for affected consumers.
Written by software from the official notice below. Not reviewed by a lawyer.
WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
Reason for recall: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere Distribution: Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The c Quantity: 3 systems Lot/code info: Catalog Number: B1018-340R UDI-DI Code: 15099590684112 Serial Numbers: 34020035 34020224 34030054
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.