← Search recalls
FDA (Device enforcement)Recalled 2025-12-16class ii

Medline Industries, LP: Medline medical procedure convenience kits labeled as: 1) LAPAROSCOPY CDS-LF, REF CDS860235L; 2) GENERAL LAPAROSCOPIC, REF CDS930223M; 3) LAP CHOLE CDS, REF CDS983678K; 4) LAPA

Hazard

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Details from the source

Reason for recall: Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use. Distribution: US Nationwide distribution. Quantity: 41432 units Lot/code info: REF CDS860235L: UDI/DI 10198459258664 (EA)40198459258665 (CS), Kit Lot Number 25EMH913; REF CDS860235L: UDI/DI 10198459258664 (EA)40198459258665 (CS), Kit Lot Number 25EMC376; REF CDS860235L: UDI/DI 10198459258664 (EA)40198459258665 (CS), Kit Lot Number 25CMJ688; REF CDS860235L: UDI/DI 10198459258664 (EA)40198459258665 (CS), Kit Lot Number 25CME476; REF CDS930223M: UDI/DI 10198459282607 (EA)40198459282608 (CS), Kit Lot Number 25EBM987; REF CDS930223M: UDI/DI 10198459282607 (EA)40198459282608 (CS), Kit Lot Number 25CBT666; REF CDS983678K: UDI/DI 10195327572242 (EA)40195327572243 (CS), Kit Lot Number 25FBP815; REF CDS983678K: UDI/DI 10195327572242 (EA)40195327572243 (CS), Kit Lot Number 25EBN002; REF CDS983678K: UDI/DI 10195327572242 (EA)40195327572243 (CS), Kit Lot Number 25CBN212; REF CDS983679K: UDI/DI 10195327641511 (EA)40195327641512 (CS), Kit Lot Number 25FBN760; REF CDS983679K: UDI/DI 10195327641511 (EA)40195327641512 (CS), Kit Lot Number 25EBD329; REF CDS983679K: UDI/DI 10195327641511 (EA)40195327641512 (CS), Kit Lot Number 25DBN040; REF CDS983679K: UDI/DI 10195327641511 (EA)40195327641512 (CS), Kit Lot Number 25DBK313; REF CDS983679K: UDI/DI 10195327641511 (EA)40195327641512 (CS), Kit Lot Number 25CBE452; REF CDS983752M: UDI/DI 10195327255855 (EA)40195327255856 (CS), Kit Lot Number 25DMH793; REF CDS983752M: UDI/DI 10195327255855 (EA)40195327255856 (CS), Kit Lot Number 25DMC405; REF CDS984016B: UDI/DI 10195327191078 (EA)40195327191079 (CS), Kit Lot Number 25EBN676; REF

UPC codes

10198459258664, 40198459258665, 10198459282607, 40198459282608, 10195327572242, 40195327572243, 10195327641511, 40195327641512, 10195327255855, 40195327255856, 10195327191078, 40195327191079, 10198459134333, 40198459134334, 10198459138959, 40198459138950, 10195327694210, 40195327694211, 10198459114564, 40198459114565, 10198459043154, 40198459043155, 10198459023774, 40198459023775, 10198459265167, 40198459265168, 10198459352614, 40198459352615, 10193489692150, 40193489692151, 10198459241611, 40198459241612, 10195327045210, 40195327045211, 10198459136382, 40198459136383, 10193489468106, 40193489468107, 10198459206641, 40198459206642

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.