DxM 1096 MicroScan WalkAway Instrument REF B1018-496
Check against the official notice below — it lists the full affected range.
Beckman Coulter Inc. is recalling 49 DxM 1096 MicroScan WalkAway Instruments worldwide that were assembled with an incorrect AC power entry connector inlet. The connector was rated for 1 ampere of current instead of the correct 10 ampere rating. Affected instruments have catalog number B1018-496, UDI-DI code 15099590690892, and serial numbers ranging from 49610039 through 49610091. In the United States, affected instruments were distributed nationwide across numerous states including Alaska, California, Florida, New York, Texas, and others.
The recall notice does not specify a remedy or recommended action for consumers who have affected instruments.
Written by software from the official notice below. Not reviewed by a lawyer.
WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
Reason for recall: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere Distribution: Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The c Quantity: 49 systems Lot/code info: Catalog Number: B1018-496 UDI-DI Code: 15099590690892 Serial Numbers: 49610039 49610040 49610041 49610042 49610043 49610044 49610045 49610046 49610047 49610048 49610050 49610051 49610052 49610053 49610054 49610055 49610056 49610057 49610058 49610059 49610060 49610062 49610063 49610064 49610065 49610066 49610067 49610068 49610069 49610070 49610071 49610072 49610073 49610074 49610075 49610078 49610079 49610080 49610081 49610082 49610083 49610084 49610085 49610086 49610087 49610088 49610089 49610090 49610091
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.