← Search recalls
FDA (Device enforcement)Recalled 2025-12-16class ii

Medline Industries, LP: Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359

Hazard

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Details from the source

Reason for recall: Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use. Distribution: US Nationwide distribution. Quantity: 96 units Lot/code info: 1) UDI/DI 10198459269196 (EA) 40198459269197 (CS), Lot Number 25EBD732; 2) UDI/DI 10198459326318 (EA) 40198459326319 (CS), Lot Number 25FBA103.

UPC codes

10198459269196, 40198459269197, 10198459326318, 40198459326319

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify