DxM 1040 MicroScan WalkAway Instrument REF B1018-440
Check against the official notice below — it lists the full affected range.
Beckman Coulter Inc. is recalling 40 DxM 1040 MicroScan WalkAway Instruments (Catalog Number B1018-440) that may have been assembled with an incorrect AC power entry connector inlet. The connector in affected units is rated for 1 ampere of current rather than the correct rating of 10 amperes. The recalled instruments were distributed worldwide and within the United States across numerous states including Alaska, California, Florida, New York, Texas, and Washington. The affected units can be identified by their serial numbers, which range from 44010026 to 44010068 (with some numbers missing from the sequence).
The recall notice does not specify a remedy or corrective action for affected consumers to take.
Written by software from the official notice below. Not reviewed by a lawyer.
WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
Reason for recall: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere Distribution: Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The c Quantity: 40 systems Lot/code info: Catalog Number: B1018-440 UDI-DI Code: 15099590690885 Serial Numbers: 44010026 44010027 44010028 44010029 44010030 44010031 44010032 44010033 44010035 44010036 44010037 44010038 44010040 44010041 44010042 44010043 44010044 44010045 44010046 44010047 44010048 44010049 44010050 44010051 44010052 44010053 44010054 44010055 44010056 44010057 44010058 44010059 44010060 44010061 44010063 44010064 44010065 44010066 44010067 44010068
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.