MicroScan WalkAway-96 plus Instrument REF B1018-284
Check against the official notice below — it lists the full affected range.
Beckman Coulter Inc. is recalling seven MicroScan WalkAway-96 plus Instruments (catalog number B1018-284) that may have been assembled with an incorrect AC power entry connector inlet. The connector inlet in these units is rated for 1 ampere of current when it should be rated for 10 amperes. The affected instruments have serial numbers 39610007 through 39610013 and were distributed worldwide and throughout numerous U.S. states.
The recall notice does not specify a remedy for affected consumers.
Written by software from the official notice below. Not reviewed by a lawyer.
WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
Reason for recall: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere Distribution: Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The c Quantity: 7 systems Lot/code info: Catalog Number: B1018-284 UDI-DI Code: 15099590658687 Serial Numbers: 39610007 39610008 39610009 39610010 39610011 39610012 39610013
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.