← Search recalls
FDA (Device enforcement)Recalled 2022-12-12class ii

Medtronic Perfusion Systems: Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB841. Used in Cardiopulmonary bypass.

Hazard

An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

Details from the source

Reason for recall: An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit. Distribution: Worldwide distribution. US Nationwide, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Belgium, Plurinational State of Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Ethiopia, Finland, France, Germany, Ghana, Greece, Guatemala, Hond Quantity: 4,462 units Lot/code info: Model CB841; Serial Number Range: 8111483548 - 8113999999; GTIN Number: 00763000225452 (Lot Numbers: 13394614, 13395228, 223819363, 223952969, 224315522, 224448403, 224772847, 224772848, 224772849); GTIN Number: 00763000225469 (Lot Numbers: 13392615, 13394868, 223476442, 223755429, 223796291, 223834097, 223834098, 224174020, 224188890, 224365018, 224419842, 224816179, 225222080, 225273494, 225273495, 225331432, 225332504, 225332505, 225332506, 225332508, 225661147)

UPC codes

00763000225452, 00763000225469

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify