← Search recalls
FDA (Device enforcement)Recalled 2024-12-19class ii

Ortho-Clinical Diagnostics, INc.: Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Model/Catalog Number: 6802783 (New), 6802914 (Certified/Refurbished), 6904000 (China) Software Ve

Hazard

The VITROS 3600 Immunodiagnostic System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.

Details from the source

Reason for recall: The VITROS 3600 Immunodiagnostic System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument. Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom. Quantity: 981 units (99 US, 882 OUS) Lot/code info: Lot Code: VITROS 3600 Immunodiagnostic System Product Code: 6802783 (New) UDI: 10758750002979 VITROS 3600 Immunodiagnostic System (Certified/Refurbished) Product Code: 6802914 UDI: 10758750002979 VITROS 3600 Immunodiagnostic System (China) Product Code: 6904000 UDI: 10758750035779 Range of installed serial numbers:36000004-36001776 Manufacture date range: June 2008 to present Distribution date range; June 2008 to present Total systems currently installed in the US: 99 Total systems currently installed outside the US: 882 Note1: The catalogue number 6802783 is used to denote the VITROS 3600 Immunodiagnostic System. Catalogue number 6802914 is used to denote the VITROS 3600 Immunodiagnostic System that have been refurbished, these systems can also be referred to as certified. A VITROS 3600 Immunodiagnostic System Certified/Refurbished maintains its original serial number through the refurbishment/certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with

UPC codes

10758750002979, 10758750035779

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify