Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
Reason for recall: Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema. Distribution: US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom Quantity: 3133 Lot/code info: UDI-DI/Lots: 05050474700901/60400378, 60401981, 60401982
05050474700901
Official record: FDA (Device enforcement) ↗