There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below). The Access Ultrasensitive Insulin APF applies an incorrect conversion factor when converting from system default units of ¿IU/mL to the International System of Units (SI units) of pmol/L.
Reason for recall: There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below). The Access Ultrasensitive Insulin APF applies an incorrect conversion factor when converting from system default units of ¿IU/mL to the International System of Units (SI units) of pmol/L. Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Croatia, Czech Republic, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom. Quantity: 19,125 reagent packs Lot/code info: UDI/DI 15099590230623, Lot Numbers: 125533, 125589, 233656, 233919, 234055, 234339, 234388