FDA (Device enforcement)Recalled 2022-02-23class ii

Siemens Medical Solutions USA, Inc: Artis Q.zen biplane, Model Number 10848355

Is my product affected?

Artis Q.zen biplane, Model Number 10848355

  • UPC 4056869010038

Check against the official notice below — it lists the full affected range.

What happened

Siemens Medical Solutions USA, Inc. is recalling 69 Artis Q.zen biplane X-ray systems (Model Number 10848355) distributed nationwide. If the coolant level in the cooling circuit drops below a certain level, the X-ray tube may not be sufficiently cooled. When this occurs, the system displays "TUBE HOT, have a break," and several minutes later blocks X-ray functionality and displays "NO XRAY: TUBE TOO HOT" to prevent further equipment damage.

The affected systems have specific serial numbers listed in the recall notice, identified by UDI 4056869010038. The source does not specify a remedy for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Details from the source

Reason for recall: If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT". Distribution: US Nationwide. Quantity: 69 systems Lot/code info: UDI: 4056869010038; Serial Numbers: 123225 123245 123026 123043 123031 123053 123202 123050 123248 123008 123255 123252 123259 123035 123211 123247 123011 123003 123032 123208 123209 123231 123246 123254 123025 123036 123220 123221 123222 123206 123207 123268 123240 123262 123265 123219 123505 123213 123250 123251 123022 123229 123205 123502 123201 123023 123024 123224 123232 123233 123234 123501 123027 123049 123249 123055 123056 123037 123264 123010 123046 123052 123200 123015 123016 123244 123212 Additional Serial Numbers as of 10/4/23: 123216 123239

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.