Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Reason for recall: Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention. Distribution: US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT. Quantity: 24 Lot/code info: Lot Code: Model No 815609360 ; UDI-DI (01)00887868144493(17)340820(10)66717603 ; Lot Number 66717603 Model No 815609360 ; UDI-DI (01)00887868144493(17)340905(10)66892902 ; Lot Number 66892902 Model No 815609360 ; UDI-DI (01)00887868144493(17)341030(10)67031550 ; Lot Number 67031550
00887868144493
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.