Artis zee floor MN, Model Number 10094142
Check against the official notice below — it lists the full affected range.
Siemens Medical Solutions USA, Inc. is recalling 24 Artis zee floor MN X-ray systems (Model Number 10094142) distributed nationwide. The systems have a coolant leak hazard in the cooling circuit that, if the coolant level drops below a certain threshold, prevents adequate cooling of the X-ray tube. When this occurs, the system displays a "TUBE HOT, have a break" message, and several minutes later blocks X-ray function and displays "NO XRAY: TUBE TOO HOT" to prevent further equipment damage.
The affected systems can be identified by their serial numbers: 140421, 140115, 140101, 140205, 140213, 140204, 140209, 140219, 140224, 140218, 140226, 140221, 140220, 140206, 140105, 140423, 140406, 140422, 140102, 140214, 140109, 140202, 140424, and 140114. The record does not specify a remedy or recommended action for affected consumers.
Written by software from the official notice below. Not reviewed by a lawyer.
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Reason for recall: If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT". Distribution: US Nationwide. Quantity: 24 systems Lot/code info: UDI: 4056869010090; Serial Numbers: 140421 140115 140101 140205 140213 140204 140209 140219 140224 140218 140226 140221 140220 140206 140105 140423 140406 140422 140102 140214 140109 140202 140424 140114
Official record: FDA (Device enforcement) ↗