FDA (Device enforcement)Recalled 2022-11-18class ii

Ventana Medical Systems, Inc.: VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 08008540001 VENTANA PD-L1 (SP142) IHC A

Hazard

Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result.

Details from the source

Reason for recall: Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result. Distribution: Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO,CT, DC, DE, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, and WI. The countries of Canada, Austria, Belgium, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Quantity: 8,927 kits Lot/code info: UDI-DI Code: 04015630967155 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 F30596 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 G27114 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 G31438 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H03958 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H13049 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H35110 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J12769 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J18775 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J25361 UDI-DI Code: 07613336106709 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 F31427 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G06715 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G16831 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G24564 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H00001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H07164 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H15132 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H18578 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H23050 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J00572 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J01568 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J14244 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001

UPC codes

04015630967155, 07613336106709, 04015630976591

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify