FDA (Device enforcement)Recalled 2014-11-10closedclass ii

DePuy Orthopaedics, Inc.: Femoral Impactor ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee replacement procedures. The three affected impactors ar

Is my product affected?

Femoral Impactor ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee replacement procedures. The three affected impactors are used at various times within the surgical flow, each impactor is used to impact and drive home its relevant mating part: Fix

  • UPC 10603295130222

Check against the official notice below — it lists the full affected range.

What happened

DePuy Orthopaedics is recalling ATTUNE INTUITION Femoral Impactors (catalog numbers 254401003, 254401004, and 254401006), which are reusable instruments used during knee replacement procedures. These impactors have been found to break and produce small pieces. If a fracture occurs during surgery but goes unnoticed, the small fractured pieces could potentially remain inside the patient's body.

A total of 4,661 affected impactors were distributed worldwide, including throughout the United States (except Alaska and Rhode Island) and to countries including Canada, Australia, Japan, South Korea, and the United Kingdom, among others. The recall record does not specify what affected consumers should do in response to this recall.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.

Details from the source

Reason for recall: ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient. Distribution: Worldwide Distribution: US (nationwide) including P.R. except AK and RI; and Internationally to: Chile, Canada, Australia, Austria, Belgium, China, France, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, United Arab Emirates (UAE), and United Kingdom (UK). Quantity: 4661 Lot/code info: CATALOG NO. : ATTUNE INTUITION Impactors Femoral Impactor, (254401006) Barcode GTIN Number: Femoral Impactor, 10603295130222 ALL LOTS

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify