Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe.
Reason for recall: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe. Distribution: US Nationwide distribution in the state of CA. Quantity: 15 units Lot/code info: UDI/DI 10193489621532 (each) 40193489621533 (case), Lot Number 25FMJ081
10193489621532, 40193489621533
Official record: FDA (Device enforcement) ↗