← Search recalls
FDA (Device enforcement)Recalled 2025-10-28class ii

Illumina, Inc.: Reagent Kits/ containing flow cells, components of NextSeq 550Dx Instrument: NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20028870/NextSeq 550Dx HO FC Cart v2.5, 75

Hazard

An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell leaks during runs, which could cause run failures, that may result in: No/insufficient data output, instrument fluidics clogs, toxic substance exposure to user.

Details from the source

Reason for recall: An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell leaks during runs, which could cause run failures, that may result in: No/insufficient data output, instrument fluidics clogs, toxic substance exposure to user. Distribution: Worldwide - US Nationwide distribution in the states of PA, NJ, CA, NC, UT, IL, MD, NY, VA, WV, SD, TX, AL, SC, TN, KY, MI and the countries of DK, SA, FR, DE, IT, CN, PL, GB, TH, GR, VN, BE, ES, IL, KR, AU, CH, BG, NL, IE, TW, TR, CZ, HU, NO, FI, JP, SG, LU, AT, AE, ZA, PT, LT, SE. Quantity: 11916 Lot/code info: NextSeq 550Dx UDI-DI: 00816270020125. Kit REF/Flow Cell REF/UDI-DI(Lot): 20028870/20031098/00816270020132(A184824-3, A185016-3, A183951-3, A182816-3, A183339-3, A182533-3, A183950-3, A184531-3, A182413-3, A184058-3, A183613-3, A183211-3, A183612-3, A183826-3, A182302-3, A183449-3, A182298-3, A182532-3, A182299-3, A183830-3, A183118-3, A182301-3, A184359-3, A184727-3); 20028871/20026365/00816270020118(A182717-3, A182051-3, A179804-3, A183343-3); 20064341/20062316/006975709330014(A183402-4, A183550-4); 20064344/20062316/006975709330038(A186185-4, A183170-4, A183404-4, A183555-4, A183557-4); 20064345/20062530/006975709330045(A186074-4, A183278-4, A183405-4, A184005-4); 20064346/20062530/006975709330052(A185054-4, A183279-4, A183409-4, A183568-4, A183569-4, A183570-4, A184776-4); 15050205(20935999, 20956354)

UPC codes

00816270020125, 00816270020132, 00816270020118

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.